Hartung, DirkScott, Jonathan B.McCabe, Steven OwenSingle, Peter2017-07-2020172017-07-202017Hartung, D., Scott, J. B., McCabe, S. O., & Single, P. (2017). MRI Safe Electrodes for Spinal Cord Stimulation (SCS). Presented at the 13th World Congress of the International Neuromodulation Society, Edinburgh, Scotland, May 27- June 1 2017.https://hdl.handle.net/10289/11206Adding decoy filars and non-therapeutic decoy electrodes to therapeutic filars can divert and dissipate a significant fraction of captured RF energy in a distributed way along the lead. Simulated worst case local SAR values have been reduced by a factor of ten for lead lengths ranging from 200 mm to 1200 mm.application/pdfen© 2017 copyright with the authors.MRI Safe Electrodes for Spinal Cord Stimulation (SCS)Conference Contribution