    Measuring the electrical impedance of mouse brain tissue
    2014
    We report on an experimental method to measure conductivity of cortical tissue. We use a pair of 5mm diameter Ag/AgCl electrodes in a Perspex sandwich device that can be brought to a distance of 400 microns apart. The apparatus is brought to uniform temperature before use. Electrical impedance of a sample is measured across the frequency range 20 Hz-2.0 MHz with an Agilent 4980A four-point impedance monitor in a shielded room. The equipment has been used to measure the conductivity of mature mouse brain cortex in vitro. Slices 400 microns in thickness are prepared on a vibratome. Slices are bathed in artificial cerebrospinal fluid (ACSF) to keep them alive. Slices are removed from the ACSF and sections of cortical tissue approximately 2 mm times 2 mm are cut with a razor blade. The sections are photographed through a calibrated microscope to allow identification of their cross-sectional areas. Excess ACSF is removed from the sample and the sections places between the electrodes. The impedance is measured across the frequency range and electrical conductivity calculated. Results show two regions of dispersion. A low frequency region is evident below approximately 10 kHz, and a high frequency dispersion above this. Results at the higher frequencies show a good fit to the Cole-Cole model of impedance of biological tissue; this model consists of resistive and non-linear capacitive elements. Physically, these elements are likely to arise due to membrane polarization and migration of ions both intra- and extra-cellularly.
  • Item type:Publication,
    In vitro electrical conductivity of seizing and non-seizing mouse brain slices at 10 kHz
    2013, Institute of Physics
    The electrical conductivity of small samples of mouse cortex (in vitro) has been measured at 10 kHz through the four-electrode method of van der Pauw. Brain slices from three mice were prepared under seizing and non-seizing conditions by changing the concentration of magnesium in the artificial cerebrospinal fluid used to maintain the tissue. These slices provided 121 square samples of cortical tissue; the conductivity of these samples was measured with an Agilent E4980A four-point impedance monitor. Of these, 73 samples were considered acceptable on the grounds of having good electrical contact between electrodes and tissue excluding outlier measurements. Results show that there is a significant difference (p = 0.03) in the conductivities of the samples under the two conditions. The seizing and non-seizing samples have mean conductivities of 0.33 and 0.36 S m⁻¹, respectively; however, these quantitative values should be used with caution as they are both subject to similar systematic uncertainties due to non-ideal temperature conditions and non-ideal placement of electrodes. We hypothesize that the difference between them, which is more robust to uncertainty, is due to the changing gap junction connectivity during seizures.
  • Item type:Publication,
    Changes in frequency of seizure-like events in stimulated cortical slices
    2010, IEEE Xplore
    Epilepsy affects nearly 3 million people in the United States alone. Given the fact that many people suffer from seizures that are intractable to pharmacological intervention, research groups are investigating the use of electrical stimulation to interact with and ameliorate symptoms of epileptic seizures. In mouse cortical slices made seizuregenic through chemical means, we applied precision controlled current pulses and measured local field potentials through a four point probe system to investigate the response of seizing tissue to electrical stimulation. We have determined that the frequency of the spontaneous seizure-like events may be modified by low amplitude, current controlled stimulation (0.5 μA). Differently from previously thought, this change in frequency is however not accompanied by any alteration of the tissue permittivity or conductivity during the inter-seizure interval.