Evaluation of detector material and radiation source position on Compton camera’s ability for multitracer imaging

Abstract

We present a study on the effects of detector material, radionuclide source and source position on the Compton camera aimed at realistic characterization of the camera's performance in multitracer imaging as it relates to brain imaging. The GEANT4 Monte Carlo simulation software was used to model the physics of radiation transport and interactions with matter. Silicon (Si) and germanium (Ge) detectors were evaluated for the scatterer, and cadmium zinc telluride (CZT) and cerium-doped lanthanum bromide (LaBr3:Ce) were considered for the absorber. Image quality analyses suggest that the use of Si as the scatterer and CZT as the absorber would be preferred. Nevertheless, two simulated Compton camera models (Si/CZT and Si/LaBr3:Ce Compton cameras) that are considered in this study demonstrated good capabilities for multitracer imaging in that four radiotracers within the nuclear medicine energy range are clearly visualized by the cameras. It is found however that beyond a range difference of about 2 cm for In-113m and F-18 radiotracers in a brain phantom, there may be a need to rotate the Compton camera for efficient brain imaging.

Citation

Uche, C. Z., Round, W. H., & Cree, M. J. (2012). Evaluation of detector material and radiation source position on Compton camera’s ability for multitracer imaging. Australasian Physical & Engineering Sciences in Medicine, 35(3), 357-364.